In vivo measures of cochlear length and insertion depth of Nucleus cochlear implant electrode arrays

ShareThis
TitleIn vivo measures of cochlear length and insertion depth of Nucleus cochlear implant electrode arrays
Publication TypeJournal Article
Year of Publication1998
AuthorsKetten, D. R., M. W. Skinner, G. Wang, M. W. Vannier, G. A. Gates, and G. J. Neely
JournalAnnals of Otology, Rhinology & Laryngology
Volume107
Issue11, Supplement 175
Pagination1-16
Call NumberDRK6802
Keywordscochlear canal length, cochlear implant, electrode array, inner ear morphometry, inner ear prosthesis
Abstract

Three-dimensional cochlear canal lengths, electrode array intracochlear insertion depths, and characteristic frequency ranges were estimated for 20 Nucleus implant recipients on the basis of in vivo computed tomography (CT) scans. Ultra-high resolution images were reconstructed from spiral CT data with O.I - mm slicing and expanded attenuation scales. Canal length estimates (mean 33.01 mm; SD 2.31) were consistent with previous findings for normal human temporal bones. Intracochlear array insertion depths estimated by 3-dimensional (3-D) spiral calculations (mean 20.19 mm; SD 2.86) and by a computerized array tracking algorithm (mean 20.36 mm; SD 2.66) were not significantly different. Estimates from surgical observations were significantly longer (mean 21.03 mm; SD 2.31) because array compressions were not detectable. Characteristic frequencies at apical electrodes estimated from Greenwood's equations ranged from 387 Hz to 2,596 Hz. The results show that significant variations in cochlear anatomy and array distribution among implant patients that may impact implant performance can be reliably detected and quantified by using in vivo high-resolution CT and 3-D reconstructions.